Product Name: Lumipulse G ¿-Amyloid 1-42-N Plasma Immunoreaction Cartridges Model/Catalog Number: 235447 Software Version: Not Applicable Product Description: Lumipulse G ¿-Amyloid 1-42-N Plasma Immunoreaction Cartridges (IRC): Lumipulse G ¿-Amyloid 1-42-N Plasma is an assay system, including a set of immunoassay reagents (Lumipulse G ¿-Amyloid 1-42-N Plasma IRC), for the quantitative measurement of ¿-Amyloid 1-42 in K2EDTA plasma specimens based on chemiluminescent enzyme immunoassay (CLEI
Product Description
Product Name: Lumipulse G ¿-Amyloid 1-42-N Plasma Immunoreaction Cartridges Model/Catalog Number: 235447 Software Version: Not Applicable Product Description: Lumipulse G ¿-Amyloid 1-42-N Plasma Immunoreaction Cartridges (IRC): Lumipulse G ¿-Amyloid 1-42-N Plasma is an assay system, including a set of immunoassay reagents (Lumipulse G ¿-Amyloid 1-42-N Plasma IRC), for the quantitative measurement of ¿-Amyloid 1-42 in K2EDTA plasma specimens based on chemiluminescent enzyme immunoassay (CLEIA) technology13 by a specific two-step immunoassay method on the LUMIPULSE G System. Lumipulse G ¿-Amyloid 1-42-N Plasma Immunoreaction Cartridges (IRC): 3 ¿ 14 Tests: 3 racks, 14 IRCs per rack, 1 test per IRC. Each IRC contains (abbreviated contents): 1. Antibody-Coated Particle Solution: (Contains anti-¿-amyloid 1-42 monoclonal antibody (mouse)-coated particles). This solution contains gelatin and turns into gel at 15¿C or lower, 2. Enzyme-Labeled Antibody Solution: (Contains alkaline phosphatase (ALP)-labeled anti-¿-amyloid 1-42 monoclonal antibodies (mouse) conjugate), 3. Assay Specific Solution: (Contains detergents in 50 mM MES buffer) Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio
Reason for Recall
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).
Details
- Recalling Firm
- Fujirebio Diagnostics, Inc.
- Units Affected
- 1423 units
- Distribution
- US Nationwide distribution in the states of Arizona, California, Florida, Indiana, Kansas, Kentucky, Maryland, Massachusetts, Minnesota, New Jersey, Pennsylvania, Texas, Washington.
- Location
- Malvern, PA
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11