Philips SureSigns VM4, VM6 and VM8 - Product Usage: patient monitors-for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics, and neonates in healthcare environments. (CHINA only) SureSigns VM4 (863063, 863085) SureSigns VM6 (863064, 863065, 863086) SureSi"gns VM8 (863066, 863068, 863087, 863088) Model Numbers: 863063,863064,863065,863066,863068,863085,863086, 453564021961,453564021971, 53564022001,453564022041, 53564022081,453564024371, 45356402
Product Description
Philips SureSigns VM4, VM6 and VM8 - Product Usage: patient monitors-for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics, and neonates in healthcare environments. (CHINA only) SureSigns VM4 (863063, 863085) SureSigns VM6 (863064, 863065, 863086) SureSi"gns VM8 (863066, 863068, 863087, 863088) Model Numbers: 863063,863064,863065,863066,863068,863085,863086, 453564021961,453564021971, 53564022001,453564022041, 53564022081,453564024371, 453564024401,453564024421 453564024441,453564024461, 453564024481
Reason for Recall
Fail to Comply with Chinese Standard YY1079-2008: Clauses 4.2.6 and 4.2.7.3. Range/accuracy of heart rate meter for pediatric mode-In pediatric mode, when the input signal rate is over 300 bpm, the indicated rate of the affected products will be lower than this upper limit
Details
- Recalling Firm
- Philips North America, LLC
- Units Affected
- 31773 units
- Distribution
- International distribution in the country of China.
- Location
- Andover, MA
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11