CF Delivery Dev Caps Bravo x5, FGS-0635/FGS 0636 - Product Usage: intended to be used for gastroesophageal pH measurement and monitoring of gastric reflux in adults and children from 4 years of age.

Reported: April 7, 2021 Initiated: February 26, 2021 #Z-1334-2021

Product Description

CF Delivery Dev Caps Bravo x5, FGS-0635/FGS 0636 - Product Usage: intended to be used for gastroesophageal pH measurement and monitoring of gastric reflux in adults and children from 4 years of age.

Reason for Recall

The capsule may fail to attach to the esophageal mucosa, which in turn may lead to the aspiration of the capsule. In the event of capsule aspiration, immediate intervention to remove the capsule is required. Potential outcomes following capsule aspiration include low oxygen saturation, intervention to retrieve the capsule, potential need to intubate the patient, extended hospital stay and a delay of treatment. Thirteen reports of capsule aspiration have been received the last 2 years.

Details

Recalling Firm
Covidien Llc
Units Affected
62,551
Distribution
Worldwide distribution - US Nationwide distribution.
Location
Mansfield, MA

Frequently Asked Questions

What product was recalled?
CF Delivery Dev Caps Bravo x5, FGS-0635/FGS 0636 - Product Usage: intended to be used for gastroesophageal pH measurement and monitoring of gastric reflux in adults and children from 4 years of age.. Recalled by Covidien Llc. Units affected: 62,551.
Why was this product recalled?
The capsule may fail to attach to the esophageal mucosa, which in turn may lead to the aspiration of the capsule. In the event of capsule aspiration, immediate intervention to remove the capsule is required. Potential outcomes following capsule aspiration include low oxygen saturation, intervention to retrieve the capsule, potential need to intubate the patient, extended hospital stay and a delay of treatment. Thirteen reports of capsule aspiration have been received the last 2 years.
Which agency issued this recall?
This recall was issued by the FDA Devices on April 7, 2021. Severity: Moderate. Recall number: Z-1334-2021.