PlainRecalls
FDA Devices Moderate Class II Ongoing

Easy Read App associated with First Response Pregnancy Test Sticks - Intended use of converting analog pregnancy test results into text pregnant or not pregnant for display on consumer devices (e.g., smart phone, iPad)

Reported: April 19, 2023 Initiated: January 25, 2023 #Z-1350-2023

Product Description

Easy Read App associated with First Response Pregnancy Test Sticks - Intended use of converting analog pregnancy test results into text pregnant or not pregnant for display on consumer devices (e.g., smart phone, iPad)

Reason for Recall

Church & Dwight First Response Pregnancy Kits was marketed with the Easy Read App without a new 510k. There is also a potential for Easy Read App to be unable to reads the test results (i.e., no results reported on the APP screen) or to misread the analog results shown on the pregnancy test stick and may lead to misinterpretation of false positive or false negative tests by the user.

Details

Recalling Firm
Church & Dwight Inc
Units Affected
780000 downloads
Distribution
US Nationwide distribution.
Location
Princeton, NJ

Frequently Asked Questions

What product was recalled?
Easy Read App associated with First Response Pregnancy Test Sticks - Intended use of converting analog pregnancy test results into text pregnant or not pregnant for display on consumer devices (e.g., smart phone, iPad). Recalled by Church & Dwight Inc. Units affected: 780000 downloads.
Why was this product recalled?
Church & Dwight First Response Pregnancy Kits was marketed with the Easy Read App without a new 510k. There is also a potential for Easy Read App to be unable to reads the test results (i.e., no results reported on the APP screen) or to misread the analog results shown on the pregnancy test stick and may lead to misinterpretation of false positive or false negative tests by the user.
Which agency issued this recall?
This recall was issued by the FDA Devices on April 19, 2023. Severity: Moderate. Recall number: Z-1350-2023.