BioPlex 2200, Anti-CCP Calibrator Set, REF 663-3200, IVD
Product Description
BioPlex 2200, Anti-CCP Calibrator Set, REF 663-3200, IVD
Reason for Recall
Bio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots. Internal testing at Bio-Rad has demonstrated that Lot No. 301481 exhibits a specificity of 94.5% (95% CI 91.8 96.4) whereas the IFU indicates a specificity of 97.8% (95% CI 96.1 98.8%). The reduction in specificity may lead to an increase in the number of false positivesBio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots. Internal testing at Bio-Rad has demonstrated that Lot No. 301481 exhibits a specificity of 94.5% (95% CI 91.8 96.4) whereas the IFU indicates a specificity of 97.8% (95% CI 96.1 98.8%). The reduction in specificity may lead to an increase in the number of false positivesBio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots. Internal testing at Bio-Rad has demonstrated that Lot No. 301481 exhibits a specificity of 94.5% (95% CI 91.8 96.4) whereas the IFU indicates a specificity of 97.8% (95% CI 96.1 98.8%). The reduction in specificity may lead to an increase in the number of false positives.There is no performance issue with the calibrator lot# 53727. They are included in the recall only because the calibrators are lot matched to the impacted reagent pack.
Details
- Recalling Firm
- Bio-Rad Laboratories, Inc.
- Units Affected
- 225 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Czech Republic, Germany, Spain, France, United Kingdom, Israel, Italy, Norway, Sweden.
- Location
- Redmond, WA
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11