Trial Cutting Guide for Triathlon Total Knee System, Triathlon PKR System, Restoris MCK Knee System, prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer. Catalog Number: I-K3834CG00 Specialty Triathlon TS Trial Cutting Guide Set I-K3834CG1L Specialty Triathlon TS Trial Cutting Guide Size 1 Left I-K3834CG1R Specialty Triathlon TS Trial Cutting Guide Size 1 Right I-K3834CG2L Specialty Triathlon TS Trial Cutting Guide Size 2 Left
Product Description
Trial Cutting Guide for Triathlon Total Knee System, Triathlon PKR System, Restoris MCK Knee System, prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer. Catalog Number: I-K3834CG00 Specialty Triathlon TS Trial Cutting Guide Set I-K3834CG1L Specialty Triathlon TS Trial Cutting Guide Size 1 Left I-K3834CG1R Specialty Triathlon TS Trial Cutting Guide Size 1 Right I-K3834CG2L Specialty Triathlon TS Trial Cutting Guide Size 2 Left I-K3834CG2R Specialty Triathlon TS Trial Cutting Guide Size 2 Right I-K3834CG3L Specialty Triathlon TS Trial Cutting Guide Size 3 Left I-K3834CG3R Specialty Triathlon TS Trial Cutting Guide Size 3 Right I-K3834CG4L Specialty Triathlon TS Trial Cutting Guide Size 4 Left I-K3834CG4R Specialty Triathlon TS Trial Cutting Guide Size 4 Right I-K3834CG5L Specialty Triathlon TS Trial Cutting Guide Size 5 Left I-K3834CG5R Specialty Triathlon TS Trial Cutting Guide Size 5 Right I-K3834CG6L Specialty Triathlon TS Trial Cutting Guide Size 6 Left I-K3834CG6R Specialty Triathlon TS Trial Cutting Guide Size 6 Right I-K3834CG7L Specialty Triathlon TS Trial Cutting Guide Size 7 Left I-K3834CG7R Specialty Triathlon TS Trial Cutting Guide Size 7 Right I-K3834CG8L Specialty Triathlon TS Trial Cutting Guide Size 8 Left I-K3834CG8R Specialty Triathlon TS Trial Cutting Guide Size 8 Right
Reason for Recall
The posterior condyle of the Specialty Triathlon TS Trial Cutting Guide(s), sizes 1 through 8, (Left and Right) has the potential to fracture at the augment cut slot. This can occur in the Trial Cutting Guide packaging during transport or during impaction of the instrument in surgery.
Details
- Recalling Firm
- Howmedica Osteonics Corp.
- Units Affected
- 113 components
- Distribution
- Distribution in US to: CA,, GA, NC, NJ, NV, OK, TX, and to Canada
- Location
- Mahwah, NJ
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11