Covidien Surgiwand" II Suction and Irrigation Device: Electrosurgical, cutting & coagulation & accessories22 Product Number/CFN: 178083, 178093, 178094
Product Description
Covidien Surgiwand" II Suction and Irrigation Device: Electrosurgical, cutting & coagulation & accessories22 Product Number/CFN: 178083, 178093, 178094
Reason for Recall
Investigations related to customer complaints received for particulate matter inside of the Surgiwand" II device tubing components identified the root cause as a damaged Y-Connector component which connects the tubing to the device. This damage was determined to generate the particulate matter observed and is created during the assembly process for connecting the tubing component to the Y-Connector.
Details
- Recalling Firm
- Covidien
- Units Affected
- 59,828 devices
- Distribution
- Domestic: AK, AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MS, NC, NE, NH, NJ, NM, NY, OH, OR, PA, SC, TX, UT, WA, WI, WV. Foreign: Canada, Japan, Taiwan, Australia, New Zealand, Indonesia, Thailand, Singapore, Malaysia, Philippines, Bahamas, Brazil, Chile, El Salvador, Guatemala, Jamaica, Mexico, Panama, Peru, Puerto Rico, Trinidad And Tobago, Austria, Belgium, Bosnia And Herzegovina, Canary Islands, Cyprus, Denmark, Equatorial Guinea, Estonia, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Kosovo, Kuwait, Lebanon, Mauritius, Netherlands, Norway, Poland, Portugal, Qatar, Romania, Russian Federation, Rwanda, Saudi Arabia, South Africa, Spain, Swaziland, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom.
- Location
- Ponce, PR
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11