RayStation Therapy Treatment Planning System, Model Numbers 2.0, 2.5, and 3.0; CLASSIFICATION NAME: System, Planning, Radiation Therapy Treatment. RayStation is a software system designed for treatment planning and analysis of radiation therapy.

Reported: June 5, 2013 Initiated: February 15, 2013 #Z-1419-2013

Product Description

RayStation Therapy Treatment Planning System, Model Numbers 2.0, 2.5, and 3.0; CLASSIFICATION NAME: System, Planning, Radiation Therapy Treatment. RayStation is a software system designed for treatment planning and analysis of radiation therapy.

Reason for Recall

Under some circumstances, imported CT, MR and PET images are offset from their true positions by one pixel in one or two directions. Users must be aware of this information to avoid incorrect dose calculations during treatment planning.

Details

Recalling Firm
RAYSEARCH LABORATORIES AB
Units Affected
Domestic: 70 units; Foreign: 27 units
Distribution
Worldwide Distribution - USA including Connecticut, Florida, Massachusetts, Missouri, New York, Ohio, and foreign distribution to(Canada, Italy, Australia, Germany, Netherlands, and Switzerland.
Location
Stockholm, N/A

Frequently Asked Questions

What product was recalled?
RayStation Therapy Treatment Planning System, Model Numbers 2.0, 2.5, and 3.0; CLASSIFICATION NAME: System, Planning, Radiation Therapy Treatment. RayStation is a software system designed for treatment planning and analysis of radiation therapy.. Recalled by RAYSEARCH LABORATORIES AB. Units affected: Domestic: 70 units; Foreign: 27 units.
Why was this product recalled?
Under some circumstances, imported CT, MR and PET images are offset from their true positions by one pixel in one or two directions. Users must be aware of this information to avoid incorrect dose calculations during treatment planning.
Which agency issued this recall?
This recall was issued by the FDA Devices on June 5, 2013. Severity: Moderate. Recall number: Z-1419-2013.