PlainRecalls
FDA Devices Moderate Class II Ongoing

BD Connecta" Plus3 White Blend OEM (sterile, bulk) Catalog No. 395214 (OUS)

Reported: July 27, 2022 Initiated: May 9, 2022 #Z-1439-2022

Product Description

BD Connecta" Plus3 White Blend OEM (sterile, bulk) Catalog No. 395214 (OUS)

Reason for Recall

May have the potential for leakage at the housing component of the stopcock, may result in delay or interruption in treatment, exposure to infusate and biohazardous material, under dosing/under infusion, contamination and/or air ingress.

Details

Units Affected
N/A
Distribution
Worldwide distribution - US Nationwide distribution in the states of AK, AR, AZ, CA, CO, FL, GA, IL, IN, KS, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NV, NY, OH, OR, PA, SD, TN, TX, UT, VA, WA, WV and the countries of Belgium, China, India, Indonesia, Japan, Korea, Nepal, Philippines, Singapore, Sri Lanka, Thailand, Vietnam.
Location
Franklin Lakes, NJ

Frequently Asked Questions

What product was recalled?
BD Connecta" Plus3 White Blend OEM (sterile, bulk) Catalog No. 395214 (OUS). Recalled by Becton Dickinson & Company. Units affected: N/A.
Why was this product recalled?
May have the potential for leakage at the housing component of the stopcock, may result in delay or interruption in treatment, exposure to infusate and biohazardous material, under dosing/under infusion, contamination and/or air ingress.
Which agency issued this recall?
This recall was issued by the FDA Devices on July 27, 2022. Severity: Moderate. Recall number: Z-1439-2022.