SOMATOM Sensation 64 Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.
Reported: May 9, 2018 Initiated: February 15, 2018 #Z-1456-2018
Product Description
SOMATOM Sensation 64 Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.
Reason for Recall
to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure
Details
- Recalling Firm
- Siemens Medical Solutions USA, Inc
- Units Affected
- 1301
- Distribution
- USA (nationally) Distribution.
- Location
- Malvern, PA
Frequently Asked Questions
What product was recalled? ▼
SOMATOM Sensation 64 Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.. Recalled by Siemens Medical Solutions USA, Inc. Units affected: 1301.
Why was this product recalled? ▼
to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on May 9, 2018. Severity: Moderate. Recall number: Z-1456-2018.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11