RaySearch Ray Station Radiation Therapy Treatment Planning System, Stand-alone software. Versions 2.0, 2.5, 3.0, 3.5 and 4.0.

Reported: April 30, 2014 Initiated: February 14, 2014 #Z-1481-2014

Product Description

RaySearch Ray Station Radiation Therapy Treatment Planning System, Stand-alone software. Versions 2.0, 2.5, 3.0, 3.5 and 4.0.

Reason for Recall

RaySearch Laboratories has recalled "RaySearch Ray Station 2.0, 2.5, 3.0, 3.5 and 4.0" software due to dose miscalculations for bolus/external/fixation support structures on CT-image.

Details

Recalling Firm
RAYSEARCH LABORATORIES AB
Units Affected
251 units (148 units domestically & 103 units internationally)
Distribution
Worldwide Distribution - USA Nationwide in the states of: AZ, CA, CT, CO, DE, FL, IL, NJ, NY, MA, MD, MS, OH, PA, TN, VA and WA. and the countries of Australia, New Zealand, Belgium, New South Wales, Republic of Korea, Italy, Germany, China, France, Switzerland, Netherlands, Spain, Austria, United Kingdom and Norway.
Location
Stockholm, N/A

Frequently Asked Questions

What product was recalled?
RaySearch Ray Station Radiation Therapy Treatment Planning System, Stand-alone software. Versions 2.0, 2.5, 3.0, 3.5 and 4.0.. Recalled by RAYSEARCH LABORATORIES AB. Units affected: 251 units (148 units domestically & 103 units internationally).
Why was this product recalled?
RaySearch Laboratories has recalled "RaySearch Ray Station 2.0, 2.5, 3.0, 3.5 and 4.0" software due to dose miscalculations for bolus/external/fixation support structures on CT-image.
Which agency issued this recall?
This recall was issued by the FDA Devices on April 30, 2014. Severity: Moderate. Recall number: Z-1481-2014.