PlainRecalls
FDA Devices Moderate Class II Ongoing

Atellica CH Iron_2 (Iron_2) is for in vitro diagnostic use in the quantitative determination of iron in human serum and (lithium heparin) plasma samples.

Reported: August 10, 2022 Initiated: June 28, 2022 #Z-1498-2022

Product Description

Atellica CH Iron_2 (Iron_2) is for in vitro diagnostic use in the quantitative determination of iron in human serum and (lithium heparin) plasma samples.

Reason for Recall

Falsely Elevated Atellica CH Microalbumin_2 (¿ALB_2) Results due to Reagent Carryover from the Iron_2 Assay. This defect occurs only when a microalbumin test is run immediately after an iron test.

Details

Units Affected
1614 units US ; 7,979 units OUS
Distribution
Worldwide distribution - US Nationwide and the countries of AR, AU, AT, BS, BH, BD, BE, BR, BG, BF, CA, CL, CO, HR, CW, CZ, DK, EC, EG, EE, FI, FR, DE, GR, GP, HK, HU, IN, IR, IQ, IE, IL, IT, JP, JO, KE, KW, LV, LY, LT, MY, MX, MA, NL, NZ, NO, OM, CN, PK, PY, PE, PH, PL, PT, QA, KR, RO, RU, SA, RS, SG, SK, ZA, ES, SE, CH, TW, TH, TR, AE, UG, GB, UY, VA, VN.
Location
Tarrytown, NY

Frequently Asked Questions

What product was recalled?
Atellica CH Iron_2 (Iron_2) is for in vitro diagnostic use in the quantitative determination of iron in human serum and (lithium heparin) plasma samples.. Recalled by Siemens Healthcare Diagnostics, Inc.. Units affected: 1614 units US ; 7,979 units OUS.
Why was this product recalled?
Falsely Elevated Atellica CH Microalbumin_2 (¿ALB_2) Results due to Reagent Carryover from the Iron_2 Assay. This defect occurs only when a microalbumin test is run immediately after an iron test.
Which agency issued this recall?