PlainRecalls
FDA Devices Moderate Class II Ongoing

Smiths Medical Medfusion Model 3500 Syringe pumps, Item Numbers: a) 3500 b) 3500-0600-00 c) 3500-0600-01 d) 3500-0600-249 e) 3500-0600-50 f) 3500-0600-51 g) 3500-0600-82 h) 3500-306 i) 3500-402 j) 3500-414 k) 3500-415 l) 3500-500 m) 3500E n) 3500VX-306 o) 3500VX-500

Reported: April 24, 2024 Initiated: December 19, 2023 #Z-1540-2024

Product Description

Smiths Medical Medfusion Model 3500 Syringe pumps, Item Numbers: a) 3500 b) 3500-0600-00 c) 3500-0600-01 d) 3500-0600-249 e) 3500-0600-50 f) 3500-0600-51 g) 3500-0600-82 h) 3500-306 i) 3500-402 j) 3500-414 k) 3500-415 l) 3500-500 m) 3500E n) 3500VX-306 o) 3500VX-500

Reason for Recall

Medfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded.

Details

Recalling Firm
Smiths Medical ASD Inc.
Units Affected
3821 pumps
Distribution
Worldwide distribution.
Location
Minneapolis, MN

Frequently Asked Questions

What product was recalled?
Smiths Medical Medfusion Model 3500 Syringe pumps, Item Numbers: a) 3500 b) 3500-0600-00 c) 3500-0600-01 d) 3500-0600-249 e) 3500-0600-50 f) 3500-0600-51 g) 3500-0600-82 h) 3500-306 i) 3500-402 j) 3500-414 k) 3500-415 l) 3500-500 m) 3500E n) 3500VX-306 o) 3500VX-500. Recalled by Smiths Medical ASD Inc.. Units affected: 3821 pumps.
Why was this product recalled?
Medfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded.
Which agency issued this recall?
This recall was issued by the FDA Devices on April 24, 2024. Severity: Moderate. Recall number: Z-1540-2024.