Smiths Medical Medfusion Model 4000 Syringe pump, Item Numbers: a) 4000-0101-249, b) 4000-0101-50, c) 4000-0101-51, d) 4000-0105-249, e) 4000-0105-50, f) 4000-0105-51, g) 4000-0105-78, h) 4000-0106-00, i) 4000-0106-01, j) 4000-0106-231,
Product Description
Smiths Medical Medfusion Model 4000 Syringe pump, Item Numbers: a) 4000-0101-249, b) 4000-0101-50, c) 4000-0101-51, d) 4000-0105-249, e) 4000-0105-50, f) 4000-0105-51, g) 4000-0105-78, h) 4000-0106-00, i) 4000-0106-01, j) 4000-0106-231,
Reason for Recall
Medfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded.
Details
- Recalling Firm
- Smiths Medical ASD Inc.
- Units Affected
- 7160 pumps
- Distribution
- Worldwide distribution.
- Location
- Minneapolis, MN
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11