DxI 9000 Access Immunoassay Analyzer, REF C11137, vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids.

Reported: September 11, 2024 Initiated: March 5, 2024 #Z-1546-2024

Product Description

DxI 9000 Access Immunoassay Analyzer, REF C11137, vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids.

Reason for Recall

Beckman Coulter manufacturing discovered cracks around a threaded insert on the bottom of the Internal Wash Buffer Tank that is used within the Wash Buffer Supply Drawer on the DxI 9000.This can result in leaking, leading to delays in reporting patient test results, exposure to physical and chemical hazards, and damage to equipment or the environment.

Details

Recalling Firm
Beckman Coulter, Inc.
Units Affected
17 units
Distribution
Worldwide - US Nationwide distribution in the state of New Hampshire and the countries of Belgium, Egypt, Italy, Poland, Spain.
Location
Chaska, MN

Frequently Asked Questions

What product was recalled?
DxI 9000 Access Immunoassay Analyzer, REF C11137, vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids.. Recalled by Beckman Coulter, Inc.. Units affected: 17 units.
Why was this product recalled?
Beckman Coulter manufacturing discovered cracks around a threaded insert on the bottom of the Internal Wash Buffer Tank that is used within the Wash Buffer Supply Drawer on the DxI 9000.This can result in leaking, leading to delays in reporting patient test results, exposure to physical and chemical hazards, and damage to equipment or the environment.
Which agency issued this recall?
This recall was issued by the FDA Devices on September 11, 2024. Severity: Moderate. Recall number: Z-1546-2024.