PlainRecalls
FDA Devices Moderate Class II Terminated

Panorama Patient Monitoring Network Mindray DS USA The Panorama Patient Monitoring can view recal time, store, print, graph and trend patient clinical and demographic data. The Panoram Patient Monitoring Network can sent independent alarm limits for data send by the bedside monitor.

Reported: May 4, 2016 Initiated: March 9, 2016 #Z-1575-2016

Product Description

Panorama Patient Monitoring Network Mindray DS USA The Panorama Patient Monitoring can view recal time, store, print, graph and trend patient clinical and demographic data. The Panoram Patient Monitoring Network can sent independent alarm limits for data send by the bedside monitor.

Reason for Recall

Panorama Central Station including the work station View Station, View Station and the eGateway will revert to the year 2000 when Daylight Savings Time (DST) occurs on March 13, 2016, or under various conditions.

Details

Units Affected
1447 units US, 45 units OUS
Distribution
Nationwide Distribution
Location
Mahwah, NJ

Frequently Asked Questions

What product was recalled?
Panorama Patient Monitoring Network Mindray DS USA The Panorama Patient Monitoring can view recal time, store, print, graph and trend patient clinical and demographic data. The Panoram Patient Monitoring Network can sent independent alarm limits for data send by the bedside monitor.. Recalled by Mindray DS USA, Inc. dba Mindray North America. Units affected: 1447 units US, 45 units OUS.
Why was this product recalled?
Panorama Central Station including the work station View Station, View Station and the eGateway will revert to the year 2000 when Daylight Savings Time (DST) occurs on March 13, 2016, or under various conditions.
Which agency issued this recall?
This recall was issued by the FDA Devices on May 4, 2016. Severity: Moderate. Recall number: Z-1575-2016.