Product Description

iQ200 Series Urine Microscopy Analyzer: a) iQ 200 Select, Part Numbers C10684 and 700-3345; b) iQ 200 Elite, Part Numbers C10683 and 700-3375; in-vitro diagnostic device, urine analysis, c) iQ200 Select 1500, Part Number 700-3347, d) iQ200 Sprint, Part Number 700-3325

Reason for Recall

Beckman Coulter has become aware of an intermittent issue in which the optional flag Previous Sample Had Sperm was enabled but not displayed so that a carryover event could have been investigated.

Details

Recalling Firm
Beckman Coulter, Inc.
Units Affected
1068 units
Distribution
Worldwide Distribution
Location
Miami, FL

Frequently Asked Questions

What product was recalled?
iQ200 Series Urine Microscopy Analyzer: a) iQ 200 Select, Part Numbers C10684 and 700-3345; b) iQ 200 Elite, Part Numbers C10683 and 700-3375; in-vitro diagnostic device, urine analysis, c) iQ200 Select 1500, Part Number 700-3347, d) iQ200 Sprint, Part Number 700-3325. Recalled by Beckman Coulter, Inc.. Units affected: 1068 units.
Why was this product recalled?
Beckman Coulter has become aware of an intermittent issue in which the optional flag Previous Sample Had Sperm was enabled but not displayed so that a carryover event could have been investigated.
Which agency issued this recall?
This recall was issued by the FDA Devices on May 24, 2023. Severity: Moderate. Recall number: Z-1596-2023.