PlainRecalls
FDA Devices Moderate Class II Terminated

MAGNETOM Aera/Skyra Systems, whole body magnetic resonance imaging systems, Model numbers 10432914, 10432915. Docking station top cover (part no# 10591430) of the removable patient table.

Reported: May 11, 2016 Initiated: April 13, 2016 #Z-1598-2016

Product Description

MAGNETOM Aera/Skyra Systems, whole body magnetic resonance imaging systems, Model numbers 10432914, 10432915. Docking station top cover (part no# 10591430) of the removable patient table.

Reason for Recall

The docking station top cover of the removable patient table may contain a magnetic metal plate. When docking/undocking the table, the magnetic top cover may become attracted to the magnet.

Details

Units Affected
8 systems
Distribution
US nationwide distribution to OH, FL, TN, MN, OH, NY, VA, and MN.
Location
Malvern, PA

Frequently Asked Questions

What product was recalled?
MAGNETOM Aera/Skyra Systems, whole body magnetic resonance imaging systems, Model numbers 10432914, 10432915. Docking station top cover (part no# 10591430) of the removable patient table.. Recalled by Siemens Medical Solutions USA, Inc. Units affected: 8 systems.
Why was this product recalled?
The docking station top cover of the removable patient table may contain a magnetic metal plate. When docking/undocking the table, the magnetic top cover may become attracted to the magnet.
Which agency issued this recall?
This recall was issued by the FDA Devices on May 11, 2016. Severity: Moderate. Recall number: Z-1598-2016.