FDA Devices Moderate Class II Ongoing

Medtronic REF: MMT-7020B, Guardian Sensor (3) for the following packaging: GS3 1PK 2L US/GTIN: 00763000140779 GS3 1PK 2L US/ GTIN: 00763000337049 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes.

Reported: May 26, 2021 Initiated: February 26, 2021 #Z-1602-2021

Product Description

Medtronic REF: MMT-7020B, Guardian Sensor (3) for the following packaging: GS3 1PK 2L US/GTIN: 00763000140779 GS3 1PK 2L US/ GTIN: 00763000337049 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes.

Reason for Recall

Use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings.

Details

Recalling Firm
Medtronic Inc.
Units Affected
302,667 units
Distribution
Worldwide distribution.
Location
Northridge, CA

Frequently Asked Questions

What product was recalled?
Medtronic REF: MMT-7020B, Guardian Sensor (3) for the following packaging: GS3 1PK 2L US/GTIN: 00763000140779 GS3 1PK 2L US/ GTIN: 00763000337049 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes.. Recalled by Medtronic Inc.. Units affected: 302,667 units.
Why was this product recalled?
Use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings.
Which agency issued this recall?
This recall was issued by the FDA Devices on May 26, 2021. Severity: Moderate. Recall number: Z-1602-2021.