Mariner RDX Polyaxial Head, REF: MX1-000020; Mariner RDX Extended Polyaxial Head, REF: MX1-000030; Mariner RDX Deformity Head, REF: MX1-000040; Mariner RDX Extended Deformity Head, REF: MX1-000050; Mariner RDX Trauma Head, REF: MX1-000060; Mariner RDX Extended Trauma Head. REF: MX1-000070, part of Mariner RDX Alpha Instruments and Implants, REF: AMRDX and Mariner RDX TD Alpha Implants, REF: AMRDXTD, which are pedicle screw systems.
Product Description
Mariner RDX Polyaxial Head, REF: MX1-000020; Mariner RDX Extended Polyaxial Head, REF: MX1-000030; Mariner RDX Deformity Head, REF: MX1-000040; Mariner RDX Extended Deformity Head, REF: MX1-000050; Mariner RDX Trauma Head, REF: MX1-000060; Mariner RDX Extended Trauma Head. REF: MX1-000070, part of Mariner RDX Alpha Instruments and Implants, REF: AMRDX and Mariner RDX TD Alpha Implants, REF: AMRDXTD, which are pedicle screw systems.
Reason for Recall
Screw heads, of pedicle screw systems, have thin lead-in thread that pose a risk of intraoperative cross-threading within the locking cap/screw head interface.
Details
- Recalling Firm
- SEASPINE ORTHOPEDICS CORPORATION
- Units Affected
- 1579 Screw Heads
- Distribution
- US Nationwide distribution in the states of MI, GA, VA, IN, FL.
- Location
- Carlsbad, CA
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11