Brand Name: Multiday lnfusor, Seven Day lnfusor and the Basal/Bolus lnfusor: Single day Infusor 2 ml/h, Single Day INFUSOR 2 ml/h System, Half Day INFUSOR SV 5 ml/h System, Two Day INFUSOR 2 ml/h System, Two day Infusor 2 ml/h, Multiday Infusor 0.5 ml/h, Multiday INFUSOR 0.5ml/h System, Seven Day Infusor 0.5 ml/h, Seven Day INFUSOR 0.5ml/h System, Basal-Bolus Infusor-15min., Basal/Bolus INFUSOR 0.5 x 2 ml/h System with 15 Minute Lockout, Basal-Bolus Infusor Device Listi
Product Description
Brand Name: Multiday lnfusor, Seven Day lnfusor and the Basal/Bolus lnfusor: Single day Infusor 2 ml/h, Single Day INFUSOR 2 ml/h System, Half Day INFUSOR SV 5 ml/h System, Two Day INFUSOR 2 ml/h System, Two day Infusor 2 ml/h, Multiday Infusor 0.5 ml/h, Multiday INFUSOR 0.5ml/h System, Seven Day Infusor 0.5 ml/h, Seven Day INFUSOR 0.5ml/h System, Basal-Bolus Infusor-15min., Basal/Bolus INFUSOR 0.5 x 2 ml/h System with 15 Minute Lockout, Basal-Bolus Infusor Device Listing: D018045 The Multiday lnfusor. Seven Day lnfusor and the BasaVBolus lnfusor are indicated for patients requiring slow, continuous intravenous, intra-arterial, subcutaneous or epidural administration of medications.
Reason for Recall
Due to an increase in complaints for leaks at the distal male Luer and Luer cap. Baxter implemented enhancements to the blue winged cap to improve the seal between the cap and the Luer.
Details
- Recalling Firm
- Baxter Healthcare Corp.
- Units Affected
- 601,230 total
- Distribution
- Worldwide Distribution - USA (nationwide), Puerto Rico, and Internationally to Brazil, Colombia, Panama, Mexico, Singapore Japan, Australia, New Zealand, Hong Kong, India, Thailand, Taiwan, and Korea.
- Location
- Round Lake, IL
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11