MAGEC 2 Rod, 5.0mm 90mm Standard; Version/Model: MC2-5090S;

Reported: April 30, 2025 Initiated: April 3, 2025 #Z-1657-2025

Product Description

MAGEC 2 Rod, 5.0mm 90mm Standard; Version/Model: MC2-5090S;

Reason for Recall

Devices for the affected lot were not assembled according to product specifications. The distraction rod outer diameter was assembled with a 5.5mm rod versus the required 5.0mm.

Details

Recalling Firm
Globus Medical, Inc.
Units Affected
1 unit
Distribution
US Nationwide distribution in the state of AR.
Location
West Norriton, PA

Frequently Asked Questions

What product was recalled?
MAGEC 2 Rod, 5.0mm 90mm Standard; Version/Model: MC2-5090S;. Recalled by Globus Medical, Inc.. Units affected: 1 unit.
Why was this product recalled?
Devices for the affected lot were not assembled according to product specifications. The distraction rod outer diameter was assembled with a 5.5mm rod versus the required 5.0mm.