Radiometer ABL90 Series - Radiometer ABL90 FLEX Model Number 393-090 Radiometer ABL90 Series - Radiometer ABL90 FLEX PLUS Model Number 393-092 The Radiometer ABL90 Series is a portable, automated analyzer for use in the laboratory environment, near patients or point-of-care settings.
Product Description
Radiometer ABL90 Series - Radiometer ABL90 FLEX Model Number 393-090 Radiometer ABL90 Series - Radiometer ABL90 FLEX PLUS Model Number 393-092 The Radiometer ABL90 Series is a portable, automated analyzer for use in the laboratory environment, near patients or point-of-care settings.
Reason for Recall
ABL90 FLEX and ABL90 FLEX PLUS analyzers running with Radiometer Windows 10 versions RWIN10 1.0 and RWIN10 1.1 may unexpectedly freeze (the analyzer ceases to operate) during a sample measurement. The analyzer software includes a feature (watchdog) that detects a freeze and automatically restarts the software to continue operating the analyzer. However, even if a restart occurs, any sample being processed at that time will be lost, potential to lead to delay in test results.
Details
- Recalling Firm
- Radiometer Medical ApS
- Units Affected
- 6275 analyzers
- Distribution
- U.S.: AL, AR, AZ, BM CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY. O.U.S.: Canada, Afghanistan, Albania, Algeria, Argentina, Bahrain, Bangladesh, Belarus, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Chile, Colombia, Cuba, Cyprus, Dominican Republic, Ecuador, Egypt, Eritrea, Ethiopia, Greece, Guatemala, Honduras, Iraq, Iran, Israel, Jamaica, Jordan, Kenya, Kuwait, Lebanon, Libya, Lithuania, Malta, Mauritius, Moldova, Nepal, Nicaragua, Nigeria, Oman, Pakistan, Paraguay, Peru, Philippines, Puerto Rico, Qatar, Romania, Saudi Arabia, Serbia, Slovenia, Sri Lanka, Suriname, Syria, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Uganda, Ukraine, United Arab Emirates, Uruguay, Uzbekistan, Venezuela, Vietnam, Yemen, Zimbabwe, Australia, New Zealand, New Caledonia, Tokelau, Austria, Belgium, Netherlands, China, Croatia, Czech Republic, Slovakia, Denmark, Iceland, Finland, Estonia, France, Germany, Hong Kong, Hungary, India, Brunei, Italy, Japan, Kazakhstan, Malaysia, Mexico, Norway, Poland, Russia, Singapore, South Africa, Cameroon, Ivory Coast, Malawi, South Korea, Spain, Portugal, Sweden, Switzerland, Turkey, United Kingdom, Ireland
- Location
- Bronshoj, N/A
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11