PlainRecalls
FDA Devices Moderate Class II Ongoing

Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount

Reported: April 30, 2025 Initiated: March 27, 2025 #Z-1663-2025

Product Description

Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount

Reason for Recall

The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column.

Details

Recalling Firm
IHB OPERATIONS B.V.
Units Affected
112 units
Distribution
US Nationwide distribution including in the states of AL, AZ, CA, FL, IA, ID, IL, LA, MA, MD, MI, MN, NC, NH, NJ, NM, NY, PA, TN, TX, UT, VA, WI, WV.
Location
Zwolle, N/A

Frequently Asked Questions

What product was recalled?
Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount. Recalled by IHB OPERATIONS B.V.. Units affected: 112 units.
Why was this product recalled?
The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column.
Which agency issued this recall?
This recall was issued by the FDA Devices on April 30, 2025. Severity: Moderate. Recall number: Z-1663-2025.