Centaur Syphilis Assay-SYP Kit, 200 Test - Product Usage: For in vitro diagnostic use in the qualitative determination of antibodies to Treponema pallidum in human serum or plasma using the ADVIA Centaur XP/XPT System and ADVIA Centaur CP Systems, SMN:10492493
Reported: June 2, 2021 Initiated: April 21, 2021 #Z-1674-2021
Product Description
Centaur Syphilis Assay-SYP Kit, 200 Test - Product Usage: For in vitro diagnostic use in the qualitative determination of antibodies to Treponema pallidum in human serum or plasma using the ADVIA Centaur XP/XPT System and ADVIA Centaur CP Systems, SMN:10492493
Reason for Recall
Potential for the ADVIA Centaur Syphilis assay to carryover into other commercial assays. This carryover effect only occurs when the impacted assays are used immediately following testing with the Syphilis assay on the ADVIA Centaur XP/XPT or ADVIA Centaur CP systems
Details
- Recalling Firm
- Siemens Healthcare Diagnostics, Inc
- Units Affected
- 22550 units
- Distribution
- Worldwide distribution - US Nationwide distribution and the countries of Austria, Belgium, Burkina Faso, Croatia, Czech Republic, Egypt, Finland, France, Fren.Polynesia, Georgia, Germany, Germany, Greece, Guadeloupe, Italy, Kazakhstan, Latvia, Netherlands, Norway, Poland, Portugal, Qatar, Russian Fed., Serbia, Slovakia, Spain, Sweden, Switzerland, Turkey, U.A.E., Ukraine, United Kingdom.
- Location
- East Walpole, MA
Frequently Asked Questions
What product was recalled? ▼
Centaur Syphilis Assay-SYP Kit, 200 Test - Product Usage: For in vitro diagnostic use in the qualitative determination of antibodies to Treponema pallidum in human serum or plasma using the ADVIA Centaur XP/XPT System and ADVIA Centaur CP Systems, SMN:10492493. Recalled by Siemens Healthcare Diagnostics, Inc. Units affected: 22550 units.
Why was this product recalled? ▼
Potential for the ADVIA Centaur Syphilis assay to carryover into other commercial assays. This carryover effect only occurs when the impacted assays are used immediately following testing with the Syphilis assay on the ADVIA Centaur XP/XPT or ADVIA Centaur CP systems
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on June 2, 2021. Severity: Moderate. Recall number: Z-1674-2021.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11