PlainRecalls
FDA Devices Moderate Class II Terminated

PIC070 Arterial Blood Sampler PICO70 - Product Usage: Preheparinized, electrolyte-balanced, arterial blood sampler product line for collection of arterial samples for pH, blood gas, oximetry, electrolyte and metabolite analysis. Model Number/Product Description: 956-518 PICO70 w/o needle; 956-519 PICO70 w/o needle; 956-522 PICO70 22G x 25 mm; 956-525 PICO70 22G x 32 mm; 956-529 PICO70 23G x 16 mm; 956-533 PICO70 23G x 25 mm; 956-534 PICO70 23G x 32 mm; 956-547 PICO70 25G x 16 mm;

Reported: June 2, 2021 Initiated: March 26, 2021 #Z-1683-2021

Product Description

PIC070 Arterial Blood Sampler PICO70 - Product Usage: Preheparinized, electrolyte-balanced, arterial blood sampler product line for collection of arterial samples for pH, blood gas, oximetry, electrolyte and metabolite analysis. Model Number/Product Description: 956-518 PICO70 w/o needle; 956-519 PICO70 w/o needle; 956-522 PICO70 22G x 25 mm; 956-525 PICO70 22G x 32 mm; 956-529 PICO70 23G x 16 mm; 956-533 PICO70 23G x 25 mm; 956-534 PICO70 23G x 32 mm; 956-547 PICO70 25G x 16 mm;

Reason for Recall

Radiometer has become aware of an issue with PICO70 Arterial Blood Sampler. The issue relates to the sterile barrier system for the product. During internal stress testing on PICO70 Arterial Blood Samplers a breach in the sterile barrier system was observed in one sample. Risk for the patient: The described error is considered to have a remote possibility of resulting in bacterial bloodstream infection. The bloodstream infection may be asymptomatic but may also progress to sepsis or life-threatening septic shock. Immunosuppressed patients are at particular risk. Due to issues (breaches) with the sterile barrier system for the Arterial Blood Sampler.

Details

Recalling Firm
Radiometer Medical ApS
Units Affected
348,178 boxes worldwide (1,883 boxes in the U.S.)
Distribution
Worldwide distribution - U.S. Nationwide distribution in the states of CA, CO, FL, IL, LA, MA, MI, MO, MT, NJ, NY, OK, PA, TX, WA, and WI. The countries of Australia, New Zealand, Canada, Algeria, Bahrain, Belarus, Bosnia and Herzegovina, Colombia, Costa Rica, Ethiopia, Greece, Guatemala, Indonesia, Lebanon, Lithuania, Malawi, Mali, Oman, Peru, Romania, Saudi Arabia, Senegal, Uganda, United Arab Emirates, Austria, Belgium, Netherlands, Luxembourg, China, Taiwan, Croatia, Czech Republic, Slovakia, Denmark, Faroe Islands, Iceland, Finland, France, Germany, Hungary, India, Maldives, Italy, Japan, Kazakhstan, Mexico, Norway, Poland, Russia, South Africa, Spain, Sweden, Switzerland, Turkey, Georgia, and United Kingdom.
Location
Bronshoj, N/A

Frequently Asked Questions

What product was recalled?
PIC070 Arterial Blood Sampler PICO70 - Product Usage: Preheparinized, electrolyte-balanced, arterial blood sampler product line for collection of arterial samples for pH, blood gas, oximetry, electrolyte and metabolite analysis. Model Number/Product Description: 956-518 PICO70 w/o needle; 956-519 PICO70 w/o needle; 956-522 PICO70 22G x 25 mm; 956-525 PICO70 22G x 32 mm; 956-529 PICO70 23G x 16 mm; 956-533 PICO70 23G x 25 mm; 956-534 PICO70 23G x 32 mm; 956-547 PICO70 25G x 16 mm;. Recalled by Radiometer Medical ApS. Units affected: 348,178 boxes worldwide (1,883 boxes in the U.S.).
Why was this product recalled?
Radiometer has become aware of an issue with PICO70 Arterial Blood Sampler. The issue relates to the sterile barrier system for the product. During internal stress testing on PICO70 Arterial Blood Samplers a breach in the sterile barrier system was observed in one sample. Risk for the patient: The described error is considered to have a remote possibility of resulting in bacterial bloodstream infection. The bloodstream infection may be asymptomatic but may also progress to sepsis or life-threatening septic shock. Immunosuppressed patients are at particular risk. Due to issues (breaches) with the sterile barrier system for the Arterial Blood Sampler.
Which agency issued this recall?
This recall was issued by the FDA Devices on June 2, 2021. Severity: Moderate. Recall number: Z-1683-2021.