PlainRecalls
FDA Devices Moderate Class II Ongoing

PERLA TL 25D SCREW REF: TLF-DS 45 25-S TLF-DS 45 30-S TLF-DS 45 35-S TLF-DS 45 40-S TLF-DS 45 45-S TLF-DS 55 30-S TLF-DS 55 35-S TLF-DS 55 40-S TLF-DS 55 45-S TLF-DS 65 35-S TLF-DS 65 40-S TLF-DS 65 45-S TLF-DS 65 50-S The PERLA¿ TL system is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sac

Reported: May 15, 2024 Initiated: March 15, 2024 #Z-1758-2024

Product Description

PERLA TL 25D SCREW REF: TLF-DS 45 25-S TLF-DS 45 30-S TLF-DS 45 35-S TLF-DS 45 40-S TLF-DS 45 45-S TLF-DS 55 30-S TLF-DS 55 35-S TLF-DS 55 40-S TLF-DS 55 45-S TLF-DS 65 35-S TLF-DS 65 40-S TLF-DS 65 45-S TLF-DS 65 50-S The PERLA¿ TL system is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease (painful degeneration of the disc), spondylolisthesis, trauma, spinal stenosis, deformities (i.e. scoliosis, kyphosis, or lordosis), tumor and failed previous fusion (pseudarthrosis).

Reason for Recall

Due to receiving information regarding difficulties with the insertion of a setscrew inside a screw head. Issues involved, either: 1. The setscrew is not tightened to torque 2. The setscrew is tightened to torque but position askew

Details

Recalling Firm
SPINEART SA
Units Affected
242 systems
Distribution
U.S. Nationwide distribution in the states of CA, and KY. Not provided due to affected devices were manufactured O.U.S.
Location
Plan-Les-Ouates, N/A

Frequently Asked Questions

What product was recalled?
PERLA TL 25D SCREW REF: TLF-DS 45 25-S TLF-DS 45 30-S TLF-DS 45 35-S TLF-DS 45 40-S TLF-DS 45 45-S TLF-DS 55 30-S TLF-DS 55 35-S TLF-DS 55 40-S TLF-DS 55 45-S TLF-DS 65 35-S TLF-DS 65 40-S TLF-DS 65 45-S TLF-DS 65 50-S The PERLA¿ TL system is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease (painful degeneration of the disc), spondylolisthesis, trauma, spinal stenosis, deformities (i.e. scoliosis, kyphosis, or lordosis), tumor and failed previous fusion (pseudarthrosis).. Recalled by SPINEART SA. Units affected: 242 systems.
Why was this product recalled?
Due to receiving information regarding difficulties with the insertion of a setscrew inside a screw head. Issues involved, either: 1. The setscrew is not tightened to torque 2. The setscrew is tightened to torque but position askew
Which agency issued this recall?
This recall was issued by the FDA Devices on May 15, 2024. Severity: Moderate. Recall number: Z-1758-2024.