PlainRecalls
FDA Devices Moderate Class II Ongoing

AquaC UNO H [120V, 60Hz, 1000VA, US-Version - a reverse osmosis unit intended for use with hemodialysis systems. Fresenius Part Number: 24-10VA-1 Vivonic Part Number: G02000970-US

Reported: September 28, 2022 Initiated: June 21, 2022 #Z-1769-2022

Product Description

AquaC UNO H [120V, 60Hz, 1000VA, US-Version - a reverse osmosis unit intended for use with hemodialysis systems. Fresenius Part Number: 24-10VA-1 Vivonic Part Number: G02000970-US

Reason for Recall

Internal ethemet cables are missing one or both ferrite cores therefore the electromagnetic compatibility of the device is not ensured

Details

Units Affected
24 units
Distribution
US Nationwide distribution in the states of AZ, CA, DC, FL, GA, IN, MA, NV, NY, PA, TX.
Location
Waltham, MA

Frequently Asked Questions

What product was recalled?
AquaC UNO H [120V, 60Hz, 1000VA, US-Version - a reverse osmosis unit intended for use with hemodialysis systems. Fresenius Part Number: 24-10VA-1 Vivonic Part Number: G02000970-US. Recalled by Fresenius Medical Care Holdings, Inc.. Units affected: 24 units.
Why was this product recalled?
Internal ethemet cables are missing one or both ferrite cores therefore the electromagnetic compatibility of the device is not ensured
Which agency issued this recall?
This recall was issued by the FDA Devices on September 28, 2022. Severity: Moderate. Recall number: Z-1769-2022.