Alaris Pump Infusion Set, Model Codes 10010453, 10010454, 10012144, 10015294, 10015414, 10933805, 11171447, 11404930, 11522558, 11607704, 10013361T, 2110-0500, 2126-0500, 2200-0500, 2411-0500, 2419-0007, 2420-0007, 2420-0500, 2421-0500, 2426-0007, 2426-0500, 2429-0500, 2433-0007, 2477-0000, 2477-0007 2478-0000 C24116E Infusion sets allow for the intravenous administration of fluids, medications, nutritional support and transfusion therapy through a needle or catheter inserted into the patient
Product Description
Alaris Pump Infusion Set, Model Codes 10010453, 10010454, 10012144, 10015294, 10015414, 10933805, 11171447, 11404930, 11522558, 11607704, 10013361T, 2110-0500, 2126-0500, 2200-0500, 2411-0500, 2419-0007, 2420-0007, 2420-0500, 2421-0500, 2426-0007, 2426-0500, 2429-0500, 2433-0007, 2477-0000, 2477-0007 2478-0000 C24116E Infusion sets allow for the intravenous administration of fluids, medications, nutritional support and transfusion therapy through a needle or catheter inserted into the patient s artery or vein.
Reason for Recall
An incomplete occlusion can be caused by a variation in the wall thickness of the pumping segment of the affected infusion sets used with the. The issue has the potential to lead to unintended delivery of medication when the pump module is not in running status or can result in faster than expected drug delivery flow when the pump is infusing.
Details
- Recalling Firm
- Becton Dickinson & Company
- Units Affected
- 17992 lots of Infusion Sets totaling 183,572,651 units (Updated 8/26/19)
- Distribution
- Worldwide Distribution; US (nationwide distribution) to following US states: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, IL, ID, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MP, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, VA, VT, WA, WI, WV, WY, and Puerto Rico; and countries of: Australia, Belgium, Canada, Chile, Malaysia, Mexico, Panama, Singapore and Taiwan.
- Location
- Franklin Lakes, NJ
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11