The LIFEPAK 500 Automated External Defibrillator is a semi-automatic defibrillator that uses a patented Shock Advisory System. This software algorithm analyzes the patient's electrocardiographic (ECG) rhythm and indicates whether or not it detects a shockable rhythm. The LIFEPAK 500 AED is intended for use by personnel who are authorized by a physician/medical director and is intended for use in the pre-hospital and workplace and community environments.

Reported: July 31, 2013 Initiated: June 12, 2013 #Z-1773-2013

Product Description

The LIFEPAK 500 Automated External Defibrillator is a semi-automatic defibrillator that uses a patented Shock Advisory System. This software algorithm analyzes the patient's electrocardiographic (ECG) rhythm and indicates whether or not it detects a shockable rhythm. The LIFEPAK 500 AED is intended for use by personnel who are authorized by a physician/medical director and is intended for use in the pre-hospital and workplace and community environments.

Reason for Recall

The LIFEPAK 500 Automated External Defibrillator (AED) could be susceptible to malfunction due to an assembly error. If affected, the malfunction would be the result of internal shorting on a printed circuit board assembly and could render the device inoperable.

Details

Recalling Firm
Physio Control, Inc.
Units Affected
139 total (125 units in the US and 14 units outside US)
Distribution
Worldwide distribution: US (nationwide) including states of: IL, MA, NC, ND, NY, PA, RI, TX, WA, and WI; and country of: Canada.
Location
Redmond, WA

Frequently Asked Questions

What product was recalled?
The LIFEPAK 500 Automated External Defibrillator is a semi-automatic defibrillator that uses a patented Shock Advisory System. This software algorithm analyzes the patient's electrocardiographic (ECG) rhythm and indicates whether or not it detects a shockable rhythm. The LIFEPAK 500 AED is intended for use by personnel who are authorized by a physician/medical director and is intended for use in the pre-hospital and workplace and community environments.. Recalled by Physio Control, Inc.. Units affected: 139 total (125 units in the US and 14 units outside US).
Why was this product recalled?
The LIFEPAK 500 Automated External Defibrillator (AED) could be susceptible to malfunction due to an assembly error. If affected, the malfunction would be the result of internal shorting on a printed circuit board assembly and could render the device inoperable.
Which agency issued this recall?
This recall was issued by the FDA Devices on July 31, 2013. Severity: Moderate. Recall number: Z-1773-2013.