SYNCHRON Systems Barbiturates Reagent (BARB 1 X 250); Catalog No. 475012 - Product Usage: BARB reagent, in conjunction with UniCel DxC 600/800 System(s) and SYNCHRON Systems Drugs of Abuse Testing (DAT) Urine Calibrators, is intended for the qualitative determination of barbiturates in human urine at a cutoff value of 200 ng/mL The BARB assay provides a rapid screening procedure for determining the presence of barbiturates (BARB) and its metabolites in urine.

Reported: May 6, 2020 Initiated: June 12, 2019 #Z-1799-2020

Product Description

SYNCHRON Systems Barbiturates Reagent (BARB 1 X 250); Catalog No. 475012 - Product Usage: BARB reagent, in conjunction with UniCel DxC 600/800 System(s) and SYNCHRON Systems Drugs of Abuse Testing (DAT) Urine Calibrators, is intended for the qualitative determination of barbiturates in human urine at a cutoff value of 200 ng/mL The BARB assay provides a rapid screening procedure for determining the presence of barbiturates (BARB) and its metabolites in urine.

Reason for Recall

The firm has identified that cross-reactivity testing may not meet the claims indicated in the Barbiturates Reagent Chemistry Information Sheet (CIS). The presence of certain compounds in urine may result in false positive or false negative results.

Details

Recalling Firm
Beckman Coulter Inc.
Units Affected
10,145 units
Distribution
Worldwide distribution - US Nationwide distribution and to multiple countries.
Location
Brea, CA

Frequently Asked Questions

What product was recalled?
SYNCHRON Systems Barbiturates Reagent (BARB 1 X 250); Catalog No. 475012 - Product Usage: BARB reagent, in conjunction with UniCel DxC 600/800 System(s) and SYNCHRON Systems Drugs of Abuse Testing (DAT) Urine Calibrators, is intended for the qualitative determination of barbiturates in human urine at a cutoff value of 200 ng/mL The BARB assay provides a rapid screening procedure for determining the presence of barbiturates (BARB) and its metabolites in urine.. Recalled by Beckman Coulter Inc.. Units affected: 10,145 units.
Why was this product recalled?
The firm has identified that cross-reactivity testing may not meet the claims indicated in the Barbiturates Reagent Chemistry Information Sheet (CIS). The presence of certain compounds in urine may result in false positive or false negative results.
Which agency issued this recall?
This recall was issued by the FDA Devices on May 6, 2020. Severity: Moderate. Recall number: Z-1799-2020.