VITROS (R) Chemistry Products Calibrator Kit 17, Product Code/ Catalog Number 6801701 -- For in vitro diagnostic use only --- VITROS Chemistry Products Calibrator Kit 17 is used in conjunction with VITROS Chemistry Products FS Calibrator 1 to calibrate VITROS 5,1 FS Chemistry System, the VITROS 4600 Chemistry System and the VITROS 5600 Integrated System for the quantitative measurement of C-reactive protein (CRP) using VITROS hsCRP Reagent.

Reported: June 25, 2014 Initiated: May 12, 2014 #Z-1804-2014

Product Description

VITROS (R) Chemistry Products Calibrator Kit 17, Product Code/ Catalog Number 6801701 -- For in vitro diagnostic use only --- VITROS Chemistry Products Calibrator Kit 17 is used in conjunction with VITROS Chemistry Products FS Calibrator 1 to calibrate VITROS 5,1 FS Chemistry System, the VITROS 4600 Chemistry System and the VITROS 5600 Integrated System for the quantitative measurement of C-reactive protein (CRP) using VITROS hsCRP Reagent.

Reason for Recall

Calibrator values assigned for the calibrator kit lots were released using a reference method that was undergoing validation for a new lot of European Reference Material (ERM). The validation process was not complete prior to calibrator value assignment for VITROS (R) Chemistry Products hsCRP Reagent (Product Code 6801739), Generation (GEN) 28.

Details

Units Affected
Domestic - 612 units, Foreign - 461 units
Distribution
Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, and Venezuela.
Location
Rochester, NY

Frequently Asked Questions

What product was recalled?
VITROS (R) Chemistry Products Calibrator Kit 17, Product Code/ Catalog Number 6801701 -- For in vitro diagnostic use only --- VITROS Chemistry Products Calibrator Kit 17 is used in conjunction with VITROS Chemistry Products FS Calibrator 1 to calibrate VITROS 5,1 FS Chemistry System, the VITROS 4600 Chemistry System and the VITROS 5600 Integrated System for the quantitative measurement of C-reactive protein (CRP) using VITROS hsCRP Reagent.. Recalled by Ortho-Clinical Diagnostics. Units affected: Domestic - 612 units, Foreign - 461 units.
Why was this product recalled?
Calibrator values assigned for the calibrator kit lots were released using a reference method that was undergoing validation for a new lot of European Reference Material (ERM). The validation process was not complete prior to calibrator value assignment for VITROS (R) Chemistry Products hsCRP Reagent (Product Code 6801739), Generation (GEN) 28.
Which agency issued this recall?
This recall was issued by the FDA Devices on June 25, 2014. Severity: Moderate. Recall number: Z-1804-2014.