META-TAN LAG/COMPRESSION SCREW KIT 90MM/85MM, REF: 71642690
Reported: May 28, 2025 Initiated: March 18, 2025 #Z-1806-2025
Product Description
META-TAN LAG/COMPRESSION SCREW KIT 90MM/85MM, REF: 71642690
Reason for Recall
Integrated Lag/Compression Screw Kit has a compression screw with hexagon socket that may be undersized, which causes the screw to be non-functional, which would require replacement with another adequate size screw, and exchange may exceed 30 minutes.
Details
- Recalling Firm
- Smith & Nephew, Inc.
- Units Affected
- 75
- Distribution
- Worldwide - US Nationwide distribution in the states of FL, NH, TX, CA, OH, NY, AL, IA, MA, NV, MO, WA, LA, TN, NC, KY, NM, NJ, AZ and the country of ZA.
- Location
- Memphis, TN
Frequently Asked Questions
What product was recalled? ▼
META-TAN LAG/COMPRESSION SCREW KIT 90MM/85MM, REF: 71642690. Recalled by Smith & Nephew, Inc.. Units affected: 75.
Why was this product recalled? ▼
Integrated Lag/Compression Screw Kit has a compression screw with hexagon socket that may be undersized, which causes the screw to be non-functional, which would require replacement with another adequate size screw, and exchange may exceed 30 minutes.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on May 28, 2025. Severity: Moderate. Recall number: Z-1806-2025.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11