The following systems are only affected if a Laird chiller is installed (11NC: 98960021599): Allura Xper Releases 3 to 8.1: Allura Xper FD10 (C+F). 722003; Allura Xper FD10/10 722005; Allura Xper FD20 722006; Allura Xper FD20/10 and FD20/20 722008; Allura Xper FD10 722010; Allura Xper FD10/10 722011; Allura Xper FD20 722012; Allura Xper FD20 biplane 722013; Allura Xper FD10 OR Table 722014; Allura Xper FD20 OR Table 722015; Allura Xper FD10/10 OR Table 722019; Allura Xper FD20 Biplane OR Ta
Product Description
The following systems are only affected if a Laird chiller is installed (11NC: 98960021599): Allura Xper Releases 3 to 8.1: Allura Xper FD10 (C+F). 722003; Allura Xper FD10/10 722005; Allura Xper FD20 722006; Allura Xper FD20/10 and FD20/20 722008; Allura Xper FD10 722010; Allura Xper FD10/10 722011; Allura Xper FD20 722012; Allura Xper FD20 biplane 722013; Allura Xper FD10 OR Table 722014; Allura Xper FD20 OR Table 722015; Allura Xper FD10/10 OR Table 722019; Allura Xper FD20 Biplane OR Table 722020; Allura Xper FD10 OR Table 722022; Allura Xper FD20 OR Table 722023; Allura Xper FD10/10 OR Table 722024; Allura Xper FD20 Biplane OR Table 722025; Allura CV20 722031
Reason for Recall
Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.
Details
- Recalling Firm
- Philips Electronics North America Corporation
- Units Affected
- 6992 (total distributed for all products)
- Distribution
- Nationwide, worldwide including Canada
- Location
- Andover, MA
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11