Coons Interventional Wire Guide, Guidewire; Reference Part Number/GPN: THSF-35-145-COONS G02356 THSF-38-145-COONS G02324 THSF-35-180-COONS G02621 THSF-35-80-COONS G12268

Reported: October 5, 2022 Initiated: August 16, 2022 #Z-1828-2022

Product Description

Coons Interventional Wire Guide, Guidewire; Reference Part Number/GPN: THSF-35-145-COONS G02356 THSF-38-145-COONS G02324 THSF-35-180-COONS G02621 THSF-35-80-COONS G12268

Reason for Recall

Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.

Details

Recalling Firm
Cook Incorporated
Units Affected
1,236 US
Distribution
Domestic distribution nationwide. Foreign distribution worldwide.
Location
Bloomington, IN