Roadrunner PC Wire Guide (The Firm), Guidewire Reference Part Number/GPN RPC-35-145 G06979 RPC-35-80 G09534

Reported: October 5, 2022 Initiated: August 16, 2022 #Z-1841-2022

Product Description

Roadrunner PC Wire Guide (The Firm), Guidewire Reference Part Number/GPN RPC-35-145 G06979 RPC-35-80 G09534

Reason for Recall

Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.

Details

Recalling Firm
Cook Incorporated
Units Affected
363 US; 244 OUS
Distribution
Domestic distribution nationwide. Foreign distribution worldwide.
Location
Bloomington, IN