FDA Devices Moderate Class II Ongoing

Torq-Flex Wire Guide Australian Modification, Guidewire Reference Part Number/GPN STF-18-40-AUST G07304

Reported: October 5, 2022 Initiated: August 16, 2022 #Z-1843-2022

Product Description

Torq-Flex Wire Guide Australian Modification, Guidewire Reference Part Number/GPN STF-18-40-AUST G07304

Reason for Recall

Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.

Details

Recalling Firm
Cook Incorporated
Units Affected
149 US
Distribution
Domestic distribution nationwide. Foreign distribution worldwide.
Location
Bloomington, IN