FDA Devices Low Class III Terminated

ECG Trunk Cables: REF/Catalog Number Description 2106305-001 ECG Trunk Cable, 3/5-lead, AHA, 3.6 m/12 ft. 2106305-002 ECG Trunk Cable, 3/5-lead, AHA, 1.2 m/4 ft. 2106305-003 ECG Trunk Cable, 3/5-lead, IEC, 3.6 m/12 ft. 2106305-004 ECG Trunk Cable, 3/5-lead, IEC, 1.2 m/4 ft. 2106306-001 ECG Trunk Cable, Neonatal, DIN 3-lead, AHA, 3.6 m/12 ft. 2106306-002 ECG Trunk Cable, Neonatal, DIN 3-lead, AHA, 1.2 m/4 ft. 2106306-003 ECG Trunk Cable, Neonatal, DIN 3-lead, IEC, 3.6 m/12 ft. 2106306-0

Reported: July 3, 2019 Initiated: May 10, 2019 #Z-1875-2019

Product Description

ECG Trunk Cables: REF/Catalog Number Description 2106305-001 ECG Trunk Cable, 3/5-lead, AHA, 3.6 m/12 ft. 2106305-002 ECG Trunk Cable, 3/5-lead, AHA, 1.2 m/4 ft. 2106305-003 ECG Trunk Cable, 3/5-lead, IEC, 3.6 m/12 ft. 2106305-004 ECG Trunk Cable, 3/5-lead, IEC, 1.2 m/4 ft. 2106306-001 ECG Trunk Cable, Neonatal, DIN 3-lead, AHA, 3.6 m/12 ft. 2106306-002 ECG Trunk Cable, Neonatal, DIN 3-lead, AHA, 1.2 m/4 ft. 2106306-003 ECG Trunk Cable, Neonatal, DIN 3-lead, IEC, 3.6 m/12 ft. 2106306-004 ECG Trunk Cable, Neonatal, DIN 3-lead, IEC, 1.2 m/4 ft. 2106307-001 ECG Trunk Cable, 6-lead, AHA, 3.6 m/12 ft. 2106307-002 ECG Trunk Cable, 6-lead, AHA, 1.2 m/4 ft. 2106307-003 ECG Trunk Cable, 6-lead, IEC, 3.6 m/12 ft. 2106307-004 ECG Trunk Cable, 6-lead, IEC, 1.2 m/4 ft. 2106308-001 ECG Trunk Cable, 12-lead, AHA, 3.6 m/12 ft. 2106308-002 ECG Trunk Cable, 12-lead, AHA, 1.2 m/4 ft. 2106308-003 ECG Trunk Cable, 12-lead, IEC, 3.6 m/12 ft. 2106308-004 ECG Trunk Cable, 12-lead, IEC, 1.2 m/4 ft. 2106310-001 ECG Trunk Cable, 3/5-lead w/ESU filter, AHA, 3.6 m/12 ft. 2106310-002 ECG Trunk Cable, 3/5-lead w/ESU filter, IEC, 3.6 m/12 ft. 2106311-001 ECG Trunk Cable, 3/5-lead, AHA, 6 m/20 ft. 2106311-002 ECG Trunk Cable, 3/5-lead, IEC, 6 m/20 ft. The reusable ECG trunk cable enables ECG monitoring with GE monitoring equipment requiring rectangular 11-pin connector cable.

Reason for Recall

ECG trunk cables and leadwires may reduce the amount of energy reaching the patient during defibrillation, potentially limiting success of defibrillating the patients rhythm.

Details

Recalling Firm
GE Healthcare, LLC
Units Affected
58,045 units (31,108 US; 26937 OUS) in total
Distribution
Worldwide Distribution: US (nationwide) including Puerto Rico, and countries of: Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Bermuda, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea (Republic of), Kuwait, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malta, Mauritius, Mexico, Moldova, Morocco, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Palestine (state of), Panama, Poland, Qatar, Republic of Serbia, Romania, Russia, Saudia Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tajikistan, Tanzania, Thailand, Trinidad and Tobago, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, and Vietnam.
Location
Waukesha, WI

Frequently Asked Questions

What product was recalled?
ECG Trunk Cables: REF/Catalog Number Description 2106305-001 ECG Trunk Cable, 3/5-lead, AHA, 3.6 m/12 ft. 2106305-002 ECG Trunk Cable, 3/5-lead, AHA, 1.2 m/4 ft. 2106305-003 ECG Trunk Cable, 3/5-lead, IEC, 3.6 m/12 ft. 2106305-004 ECG Trunk Cable, 3/5-lead, IEC, 1.2 m/4 ft. 2106306-001 ECG Trunk Cable, Neonatal, DIN 3-lead, AHA, 3.6 m/12 ft. 2106306-002 ECG Trunk Cable, Neonatal, DIN 3-lead, AHA, 1.2 m/4 ft. 2106306-003 ECG Trunk Cable, Neonatal, DIN 3-lead, IEC, 3.6 m/12 ft. 2106306-004 ECG Trunk Cable, Neonatal, DIN 3-lead, IEC, 1.2 m/4 ft. 2106307-001 ECG Trunk Cable, 6-lead, AHA, 3.6 m/12 ft. 2106307-002 ECG Trunk Cable, 6-lead, AHA, 1.2 m/4 ft. 2106307-003 ECG Trunk Cable, 6-lead, IEC, 3.6 m/12 ft. 2106307-004 ECG Trunk Cable, 6-lead, IEC, 1.2 m/4 ft. 2106308-001 ECG Trunk Cable, 12-lead, AHA, 3.6 m/12 ft. 2106308-002 ECG Trunk Cable, 12-lead, AHA, 1.2 m/4 ft. 2106308-003 ECG Trunk Cable, 12-lead, IEC, 3.6 m/12 ft. 2106308-004 ECG Trunk Cable, 12-lead, IEC, 1.2 m/4 ft. 2106310-001 ECG Trunk Cable, 3/5-lead w/ESU filter, AHA, 3.6 m/12 ft. 2106310-002 ECG Trunk Cable, 3/5-lead w/ESU filter, IEC, 3.6 m/12 ft. 2106311-001 ECG Trunk Cable, 3/5-lead, AHA, 6 m/20 ft. 2106311-002 ECG Trunk Cable, 3/5-lead, IEC, 6 m/20 ft. The reusable ECG trunk cable enables ECG monitoring with GE monitoring equipment requiring rectangular 11-pin connector cable.. Recalled by GE Healthcare, LLC. Units affected: 58,045 units (31,108 US; 26937 OUS) in total.
Why was this product recalled?
ECG trunk cables and leadwires may reduce the amount of energy reaching the patient during defibrillation, potentially limiting success of defibrillating the patients rhythm.
Which agency issued this recall?
This recall was issued by the FDA Devices on July 3, 2019. Severity: Low. Recall number: Z-1875-2019.