Replacement leadwires with grabber ends: REF/Catalog Number Description 2106388-005 Replacement ECG Leadwire, grabber, BRN V, AHA, 74 cm/ 29 in 2106388-101 Replacement ECG Leadwire, grabber, LL , AHA, 130 cm/ 51 in 2106388-103 Replacement ECG Leadwire, grabber, GRN RL, AHA, 130 cm/ 2106388-107 Replacement ECG Leadwire, grabber, BLK N, IEC, 130 cm/ 51 in 2106392-004 Replacement ECG Leadwire, grabber, ORG V, AHA, 74 cm/ 29 in The reusable ECG leadwire sets enable ECG monitoring with GE eq
Product Description
Replacement leadwires with grabber ends: REF/Catalog Number Description 2106388-005 Replacement ECG Leadwire, grabber, BRN V, AHA, 74 cm/ 29 in 2106388-101 Replacement ECG Leadwire, grabber, LL , AHA, 130 cm/ 51 in 2106388-103 Replacement ECG Leadwire, grabber, GRN RL, AHA, 130 cm/ 2106388-107 Replacement ECG Leadwire, grabber, BLK N, IEC, 130 cm/ 51 in 2106392-004 Replacement ECG Leadwire, grabber, ORG V, AHA, 74 cm/ 29 in The reusable ECG leadwire sets enable ECG monitoring with GE equipment requiring rectangular 11-pin ECG trunk cable. Use leadwire sets with ECG trunk cable or telemetry transmitter.
Reason for Recall
ECG trunk cables and leadwires may reduce the amount of energy reaching the patient during defibrillation, potentially limiting success of defibrillating the patients rhythm.
Details
- Recalling Firm
- GE Healthcare, LLC
- Units Affected
- 58,045 units (31,108 US; 26937 OUS) in total
- Distribution
- Worldwide Distribution: US (nationwide) including Puerto Rico, and countries of: Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Bermuda, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea (Republic of), Kuwait, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malta, Mauritius, Mexico, Moldova, Morocco, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Palestine (state of), Panama, Poland, Qatar, Republic of Serbia, Romania, Russia, Saudia Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tajikistan, Tanzania, Thailand, Trinidad and Tobago, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, and Vietnam.
- Location
- Waukesha, WI
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11