AirLife Isothermal Breathing Circuit, Infant Respiratory Circuit, Non-Heated; an Rx device for single patient use; CareFusion, 22745 Savi Ranch Parkway, Yorba Linda, CA 92887 USA, Made in Mexico Respiratory breathing circuits are used with a ventilator when mechanical ventilation is administered to a patient.

Reported: July 11, 2012 Initiated: May 29, 2012 #Z-1900-2012

Product Description

AirLife Isothermal Breathing Circuit, Infant Respiratory Circuit, Non-Heated; an Rx device for single patient use; CareFusion, 22745 Savi Ranch Parkway, Yorba Linda, CA 92887 USA, Made in Mexico Respiratory breathing circuits are used with a ventilator when mechanical ventilation is administered to a patient.

Reason for Recall

The Y adapter within the breathing circuit may spontaneously crack, causing a leak in the closed ventilation system. This leak could lead to a decrease in the tidal volume delivered to the ventilated patient.

Details

Recalling Firm
Carefusion 2200 Inc
Units Affected
N/A
Distribution
Worldwide distribution: USA (nationwide) including Puerto Rico and internationally to: Canada, Egypt, France, Germany, Great Britain, Italy, Kuwait, Mexico, Portugal, Saudi Arabia, Spain, Taiwan and United Arab Emirates.
Location
McGaw Park, IL

Frequently Asked Questions

What product was recalled?
AirLife Isothermal Breathing Circuit, Infant Respiratory Circuit, Non-Heated; an Rx device for single patient use; CareFusion, 22745 Savi Ranch Parkway, Yorba Linda, CA 92887 USA, Made in Mexico Respiratory breathing circuits are used with a ventilator when mechanical ventilation is administered to a patient.. Recalled by Carefusion 2200 Inc. Units affected: N/A.
Why was this product recalled?
The Y adapter within the breathing circuit may spontaneously crack, causing a leak in the closed ventilation system. This leak could lead to a decrease in the tidal volume delivered to the ventilated patient.
Which agency issued this recall?
This recall was issued by the FDA Devices on July 11, 2012. Severity: Critical. Recall number: Z-1900-2012.