FDA Devices Moderate Class II Terminated

Medtronic DLP Femoral Arterial Cannula, models 57414, 57417, and 57421, and DLP Femoral Cannula Replacement Kit, Models 96017 and 96021. Sterilized using Ethylene Oxide. The cannula and dilator components are polyurethane and have tapered proximal tips to aid in diameter transition from the dilator. The dilator components utilize tip indicator bands for minimum tip alignment of the component parts. The cannula body is marked for depth insertion. The components are designed to fit over eac

Reported: July 8, 2015 Initiated: June 12, 2015 #Z-1914-2015

Product Description

Medtronic DLP Femoral Arterial Cannula, models 57414, 57417, and 57421, and DLP Femoral Cannula Replacement Kit, Models 96017 and 96021. Sterilized using Ethylene Oxide. The cannula and dilator components are polyurethane and have tapered proximal tips to aid in diameter transition from the dilator. The dilator components utilize tip indicator bands for minimum tip alignment of the component parts. The cannula body is marked for depth insertion. The components are designed to fit over each other to progressively dilate the vessel during insertion. The dilator allows a 0.1-cm (0.038-in) guidewire to pass through it to aid in insertion. A soft, tapered transition fitting terminates in a 0.95-cm (3/8-in) connection site or barbed connector.

Reason for Recall

Medtronic is recalling select lots of DLP Femoral Arterial Cannula product due to recent reports in which the user had difficulty or was unable to connect the cannula to the perfusion circuit tubing.

Details

Units Affected
1304 devices
Distribution
Worldwide Distribution - US Distribution including the states of : AK, AZ, CA, CO, IN, KS, LA, MN, MT, NM, NY, OH, OR, PA, TX, UT, and WI., and to the countries of : Belgium, Brazil, Canada, Colombia, Denmark, France, Germany, Greece, Hungary, India, Ireland, Italy, Korea, Morocco, Netherlands, Poland, Qatar, Russian Federation, Saudi Arabia, Slovakia, Spain, Switzerland, Turkey, and United Kingdom.
Location
Brooklyn Park, MN

Frequently Asked Questions

What product was recalled?
Medtronic DLP Femoral Arterial Cannula, models 57414, 57417, and 57421, and DLP Femoral Cannula Replacement Kit, Models 96017 and 96021. Sterilized using Ethylene Oxide. The cannula and dilator components are polyurethane and have tapered proximal tips to aid in diameter transition from the dilator. The dilator components utilize tip indicator bands for minimum tip alignment of the component parts. The cannula body is marked for depth insertion. The components are designed to fit over each other to progressively dilate the vessel during insertion. The dilator allows a 0.1-cm (0.038-in) guidewire to pass through it to aid in insertion. A soft, tapered transition fitting terminates in a 0.95-cm (3/8-in) connection site or barbed connector.. Recalled by Medtronic Perfusion Systems. Units affected: 1304 devices.
Why was this product recalled?
Medtronic is recalling select lots of DLP Femoral Arterial Cannula product due to recent reports in which the user had difficulty or was unable to connect the cannula to the perfusion circuit tubing.
Which agency issued this recall?
This recall was issued by the FDA Devices on July 8, 2015. Severity: Moderate. Recall number: Z-1914-2015.