PlainRecalls
FDA Devices Moderate Class II Terminated

iChem¿VELOCITY Urine Chemistry System-

Reported: May 20, 2020 Initiated: March 30, 2020 #Z-1944-2020

Product Description

iChem¿VELOCITY Urine Chemistry System-

Reason for Recall

A defect in the SATA power adapter cable within the instruments computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or melting of the SATA adapter cable within the computers metal enclosure, smoke emanating from PC6, flames contained within PC6, or an inability to use the devices CD-ROM.

Details

Recalling Firm
Beckman Coulter Inc.
Units Affected
1557 units
Distribution
US Distribution: Nationwide Foreign Distribution: Worldwide
Location
Brea, CA

Frequently Asked Questions

What product was recalled?
iChem¿VELOCITY Urine Chemistry System-. Recalled by Beckman Coulter Inc.. Units affected: 1557 units.
Why was this product recalled?
A defect in the SATA power adapter cable within the instruments computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or melting of the SATA adapter cable within the computers metal enclosure, smoke emanating from PC6, flames contained within PC6, or an inability to use the devices CD-ROM.
Which agency issued this recall?
This recall was issued by the FDA Devices on May 20, 2020. Severity: Moderate. Recall number: Z-1944-2020.