FDA Devices Moderate Class II Terminated

Access 2 Immunoassay System, Remanufactured, Catalog No. 386220 is an invitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analytes concentrations found in human body fluids.

Reported: June 22, 2016 Initiated: May 17, 2016 #Z-1956-2016

Product Description

Access 2 Immunoassay System, Remanufactured, Catalog No. 386220 is an invitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analytes concentrations found in human body fluids.

Reason for Recall

Due to the lack of updated thermal components, these instruments may be more susceptible to the internal case temperature exceeding the 18 to 36 degrees Celsius operating range, causing SYS flags. The instrument will not report results until the internal case temperature returns to the operating range. This situation may cause a delay in the reporting of patient results.

Details

Recalling Firm
Beckman Coulter Inc.
Units Affected
4 units total (1 unit in US)
Distribution
US, Morocco, Venezuela
Location
Brea, CA

Frequently Asked Questions

What product was recalled?
Access 2 Immunoassay System, Remanufactured, Catalog No. 386220 is an invitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analytes concentrations found in human body fluids.. Recalled by Beckman Coulter Inc.. Units affected: 4 units total (1 unit in US).
Why was this product recalled?
Due to the lack of updated thermal components, these instruments may be more susceptible to the internal case temperature exceeding the 18 to 36 degrees Celsius operating range, causing SYS flags. The instrument will not report results until the internal case temperature returns to the operating range. This situation may cause a delay in the reporting of patient results.
Which agency issued this recall?
This recall was issued by the FDA Devices on June 22, 2016. Severity: Moderate. Recall number: Z-1956-2016.