00597001101 Femur CR PRECOAT FEM COMP SIZE AML, Rx, Sterile; 00597001102 Femur CR PRECOAT FEM COMP SIZE AMR, Rx, Sterile; 00597001201 Femur CR PRECOAT FEM COMP SIZE BML, Rx, Sterile; 00597001202 Femur CR PRECOAT FEM COMP SIZE BMR, Rx, Sterile. Used in total knee arthroplasty and indicated for patients with severe pain.
Product Description
00597001101 Femur CR PRECOAT FEM COMP SIZE AML, Rx, Sterile; 00597001102 Femur CR PRECOAT FEM COMP SIZE AMR, Rx, Sterile; 00597001201 Femur CR PRECOAT FEM COMP SIZE BML, Rx, Sterile; 00597001202 Femur CR PRECOAT FEM COMP SIZE BMR, Rx, Sterile. Used in total knee arthroplasty and indicated for patients with severe pain.
Reason for Recall
Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d
Details
- Recalling Firm
- Zimmer, Inc.
- Units Affected
- 192,355 all devices
- Distribution
- Worldwide Distribution-USA (nationwide) and the country of Japan.
- Location
- Warsaw, IN
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11