INGENIO DR EL (pacemakers), Model Numbers: a) J174 b) J177 c) K174 (added 7/15/2021) d) K184 (added 7/15/2021) e) K187 (added 7/15/2021)

Reported: July 21, 2021 Initiated: June 3, 2021 #Z-2016-2021

Product Description

INGENIO DR EL (pacemakers), Model Numbers: a) J174 b) J177 c) K174 (added 7/15/2021) d) K184 (added 7/15/2021) e) K187 (added 7/15/2021)

Reason for Recall

The dual chamber INGENIO family pacemakers or cardiac resynchronization therapy pacemakers (CRT-Ps) may initiate Safety Mode later in device life (i.e., prior to reaching the Explant battery indicator) when the device's battery exhibits high internal impedance. This latent battery condition puts a device at risk for system resets to occur due to temporary high-power consumption related to telemetry attempts and subsequent reversion to Safety Mode to maintain back-up pacing. Although therapy is still provided when a device is in Safety Mode, replacement is required. - Only dual chamber INGENIO family pacemakers and CRT-Ps built with the Extended Life (EL) battery are included within this advisory population

Details

Units Affected
40,224 units
Distribution
Worldwide distribution.
Location
Saint Paul, MN

Frequently Asked Questions

What product was recalled?
INGENIO DR EL (pacemakers), Model Numbers: a) J174 b) J177 c) K174 (added 7/15/2021) d) K184 (added 7/15/2021) e) K187 (added 7/15/2021). Recalled by Boston Scientific Corporation. Units affected: 40,224 units.
Why was this product recalled?
The dual chamber INGENIO family pacemakers or cardiac resynchronization therapy pacemakers (CRT-Ps) may initiate Safety Mode later in device life (i.e., prior to reaching the Explant battery indicator) when the device's battery exhibits high internal impedance. This latent battery condition puts a device at risk for system resets to occur due to temporary high-power consumption related to telemetry attempts and subsequent reversion to Safety Mode to maintain back-up pacing. Although therapy is still provided when a device is in Safety Mode, replacement is required. - Only dual chamber INGENIO family pacemakers and CRT-Ps built with the Extended Life (EL) battery are included within this advisory population
Which agency issued this recall?
This recall was issued by the FDA Devices on July 21, 2021. Severity: Critical. Recall number: Z-2016-2021.