SmartSite Low Sorbing Infusion Set, Model Codes: 10014855A, 10015861A, 10061661A 14035457 14035485 14046894 14035458
Reported: July 16, 2014 Initiated: June 17, 2014 #Z-2025-2014
Product Description
SmartSite Low Sorbing Infusion Set, Model Codes: 10014855A, 10015861A, 10061661A 14035457 14035485 14046894 14035458
Reason for Recall
CareFusion is recalling the SmartSite Low Sorbing Infusion Set because they were assembled with a drip chamber that consist of a 15 micron filter. The Directions For Use did not indicate that the drip chamber has a 15 micron filter and it does not provide the instruction that the set should not be used with blood administration due to the potential for occluding the 15 micron filter in the drip c
Details
- Recalling Firm
- CareFusion 303, Inc.
- Units Affected
- 5,600 units
- Distribution
- Distributed in the states of FL, NM, WA, WI, and TX.
- Location
- San Diego, CA
Frequently Asked Questions
What product was recalled? ▼
SmartSite Low Sorbing Infusion Set, Model Codes: 10014855A, 10015861A, 10061661A 14035457 14035485 14046894 14035458. Recalled by CareFusion 303, Inc.. Units affected: 5,600 units.
Why was this product recalled? ▼
CareFusion is recalling the SmartSite Low Sorbing Infusion Set because they were assembled with a drip chamber that consist of a 15 micron filter. The Directions For Use did not indicate that the drip chamber has a 15 micron filter and it does not provide the instruction that the set should not be used with blood administration due to the potential for occluding the 15 micron filter in the drip c
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on July 16, 2014. Severity: Moderate. Recall number: Z-2025-2014.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11