HeartStart MRx Monitor/Defibrillator, Model No. M3535A (861288), M3536A (861289) Product Automated external defibrillators (non-wearable) Product Usage: The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physici

Reported: June 6, 2018 Initiated: May 3, 2018 #Z-2035-2018

Product Description

HeartStart MRx Monitor/Defibrillator, Model No. M3535A (861288), M3536A (861289) Product Automated external defibrillators (non-wearable) Product Usage: The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician.

Reason for Recall

The MRx monitor/defibrillators could fail to charge because the therapy printed circuit board may have been loaded with an incorrect electronic component. Failure to charge could potentially cause therapy to be interrupted or delayed.

Details

Units Affected
5
Distribution
Worldwide Distribution to accounts in OH and NC. Foreign distribution to Lebanon.
Location
Andover, MA

Frequently Asked Questions

What product was recalled?
HeartStart MRx Monitor/Defibrillator, Model No. M3535A (861288), M3536A (861289) Product Automated external defibrillators (non-wearable) Product Usage: The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician.. Recalled by Philips Electronics North America Corporation. Units affected: 5.
Why was this product recalled?
The MRx monitor/defibrillators could fail to charge because the therapy printed circuit board may have been loaded with an incorrect electronic component. Failure to charge could potentially cause therapy to be interrupted or delayed.
Which agency issued this recall?
This recall was issued by the FDA Devices on June 6, 2018. Severity: Moderate. Recall number: Z-2035-2018.