EV3 EverFiex" Self-expanding Peripheral Stent System, Sterilized using Ethylene Oxide, Rx Only, PRB35-08-040-080. The Protege EverFiex" Self-expanding Peripheral Stent System is a self-expanding Nitinol stent system intended for permanent implantation. The self-expanding stent is made of a nickel titanium alloy (Nitinol) and comes pre-mounted on a 6 F, 0.035" over-the-wire delivery system. The stent is cut from a Nitinol tube in an open lattice design, and has tantalum radiopaque markers at th
Product Description
EV3 EverFiex" Self-expanding Peripheral Stent System, Sterilized using Ethylene Oxide, Rx Only, PRB35-08-040-080. The Protege EverFiex" Self-expanding Peripheral Stent System is a self-expanding Nitinol stent system intended for permanent implantation. The self-expanding stent is made of a nickel titanium alloy (Nitinol) and comes pre-mounted on a 6 F, 0.035" over-the-wire delivery system. The stent is cut from a Nitinol tube in an open lattice design, and has tantalum radiopaque markers at the proximal and distal ends of the stent. Upon deployment, the stent achieves its predetermined diameter and exerts a constant, gentle outward force to establish patency
Reason for Recall
Medtronic is issuing a medical device recall to return product potentially associated with an outer box mislabel issue for two Protege EverFiex" Self-expanding Peripheral Stent Systems, model PRB35-08-040- 080, from lot A164580. The size on the outer box is labeled as 8x40x80mm while potentially containing a 6x40x120mm device as indicated by both the inner pouch and device labels. This potential issue is limited to two EverFiex stent system devices from lot A164580 that were repackaged by Medtronic and distributed. This does not affect any other Medtronic product or implantable devices.
Details
- Recalling Firm
- Ev3, Inc.
- Units Affected
- 2
- Distribution
- Nationwide Distribution to LA and WI.
- Location
- Plymouth, MN
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11