*** 1) Product labeled in part: 19683-28; LIFESHIELD; MICRODRIP SOLUSET WITH 1.2 MICRON FILTER 150ML BURETTE PLUMSET, CONVERTIBLE PIN, CLAVE SECONDARY PORT, OPTION-LOK, NON-DEHP; NOMINAL LENGTH: 118 IN; PRIMING VOLUME: 14 ML; *** 2) Product labeled in part: List No. 19685-28; LifeShield; LATEX-FREE; PLUMSET, PRIMARY CONVERTIBLE PP W/ SIGHT CHAMBER, THREE WAY STOPCOCK, CLAVE INJECTION PORTS, OPTION-LOK, NON-DEHP; NOMINAL LENGTH: 136 IN (345 CM); PRIMING VOLUME: 24 ML; *** 3) Product labeled
Product Description
*** 1) Product labeled in part: 19683-28; LIFESHIELD; MICRODRIP SOLUSET WITH 1.2 MICRON FILTER 150ML BURETTE PLUMSET, CONVERTIBLE PIN, CLAVE SECONDARY PORT, OPTION-LOK, NON-DEHP; NOMINAL LENGTH: 118 IN; PRIMING VOLUME: 14 ML; *** 2) Product labeled in part: List No. 19685-28; LifeShield; LATEX-FREE; PLUMSET, PRIMARY CONVERTIBLE PP W/ SIGHT CHAMBER, THREE WAY STOPCOCK, CLAVE INJECTION PORTS, OPTION-LOK, NON-DEHP; NOMINAL LENGTH: 136 IN (345 CM); PRIMING VOLUME: 24 ML; *** 3) Product labeled in part: 19720-28; LIFESHIELD; PLUM SET, CONV P.P, DUAL CHANNEL 3 CLAVE PORTS,IN LINE CHECK VAL OPTION LOK, NON-DEHP.; NOMINAL LENGTH:145 IN (368 CM); PRIMING VOLUME:13 ML; *** 4) Product labeled in part: 19721-28; LIFESHIELD; SOLUSET 150ML BUR,CONV P.P DUAL CHANNEL 4 CLAVE PORTS IN LINE CHECK VAL OPTION LOK, NON-DEHP; NOMINAL LENGTH: 145 IN (368 CM); PRIMING VOLUME: 15 ML: *** 5) Product labeled in part: 19729-28; LIFESHIELD; CONV. DUAL CHANNEL PLUMSET WITH CLAVES, AND SECURE LOK, NON-DEHP; NOMINAL LENGTH: 96 IN (244 CM); PRIMING VOLUME: 18 ML; *** 6) Product labeled in part: 19732-28; LIFESHIELD; LIFESHIELD LATEX-FREE PLUM SET,MICRODRIP,PP CHECK VALVE AND CLAVE OPTION-LOK, NON-DEHP; NOMINAL LENGTH:104 IN (264 CM); PRIMING VOLUME: 13 ML; Product Usage: The CLAVE of the secondary port is used as an easy secondary access that allows concurrent and piggyback mode in infusion. It is normally used with a syringe or a secondary IV infusion set.
Reason for Recall
Intravenous fluid may leak at the CLAVE secondary port of Hospira PlumSet IV administration sets.
Details
- Recalling Firm
- Hospira Inc.
- Units Affected
- 3,920,932 units in total
- Distribution
- US (Nationwide) Distribution
- Location
- Lake Forest, IL
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11