Reason for Recall
A firmware error may lead to an incorrect association of test result(s) to a sample ID and therefore to an incorrect patient result. The mis-association of test results, or a delay in testing, may occur when all the conditions listed below occur within a few milliseconds timeframe of each other: - The SRM module is releasing a sample tube (Tube A) just placed into the carrier - Another sample tube (Tube B) is erroneously not diverted into the SRM module buffer - There is divert gate malfunction and a narrow timing window Only in this specific scenario, Tube A may be released by the module as Tube B due to a miscommunication between the module firmware and the Automation software without a unique error message being generated. The Automation System loses the traceability of Tube A. It manages both Tube A (incorrectly identified as Tube B) and the real Tube B according to the pending test orders (not yet performed) on Tube B.
Details
- Recalling Firm
- Siemens Healthcare Diagnostics, Inc.
- Units Affected
- 83 units
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of CA, CO, FL, IL, KY, LA, MA, MD, MI, MS, MT, NC, NH, NY, OH, PA, SC, TN, TX, VA, WA, & WV. The countries of Australia, Brazil, Canada, China, France, Germany, Greece, Italy, Portugal, Saudi Arabia, Spain, Sweden, & United Kingdom.
- Location
- Newark, DE
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11